Lab and field counts · Haematology
How to do a manual WBC differential count, and the battlement method

The tool for this guide: Open the differential counter
A manual differential is counted from a stained blood film under oil immersion, moving in a battlement pattern through the monolayer and classifying each white cell until a set number is reached. One hundred cells is the routine count. The percentages are then reported, and may be converted to absolute counts.
What the count is, and when it is done
A differential white cell count reports what proportion of the white cells present belong to each class — neutrophils, lymphocytes, monocytes, eosinophils, basophils, and any immature or abnormal forms seen. Modern analysers produce a differential automatically on most samples. The manual count exists for the cases where the analyser flags something, cannot classify what it is seeing, or where morphology itself is the question.
That is worth stating plainly, because it frames everything below: the manual differential is the method used precisely when the answer matters and the machine would not do.
Two kinds of cell on the film fall outside the five classes. Immature and abnormal white cells, such as blasts, are white cells, so they belong in the count under a class of their own. Nucleated red cells are not, so they are tallied beside the count and reported per 100 white cells, which is the figure used to correct an analyser’s white count for them. The differential counter has a key for each, and does that correction from the count you type.
The hardest call in the count is the band: band vs segmented neutrophil sets out the filament rule, the one-third rule and how often observers disagree. Once the differential is counted, the absolute neutrophil count is the white count times the segmented neutrophils and bands added together.
Where on the film to count
Not anywhere. A blood film is not uniform, and where you count changes what you count.
The count is made in the monolayer — the region where red cells lie in a single layer, touching but not overlapping. Too near the head of the film the cells are piled up and unreadable. Too near the tail and the distribution is distorted: larger cells, particularly monocytes and neutrophils, are dragged toward the edges and the feathered end, while lymphocytes are relatively over-represented in the middle. Counting in the wrong region does not add random noise; it adds a bias in a predictable direction, which is worse.
The monolayer only exists on a well-made wedge smear, and the same monolayer is where a platelet estimate is counted.
The battlement method, and why
The battlement method is a fixed path for choosing which fields of a blood film to count in a differential: a few fields along the edge of the film, a few inward at right angles, a few parallel to the edge again, then back out to the edge, so the track looks like the top of a castle wall. Every field on the path is counted completely before moving on.
The reason is the bias above. A straight track along the film samples one band of it and inherits whatever distortion that band carries. The battlement deliberately crosses between the edge and the centre repeatedly, so that over-represented and under-represented regions are sampled in turn and their biases partly cancel. The pattern is not ritual, it is a sampling correction.
Where it comes from: the 1940 comparison
The method was set against its rivals by R. G. Scott MacGregor, W. Richards and G. L. Loh in “The differential leucocyte count”, The Journal of Pathology and Bacteriology 51(3), 1940, pages 337–368 (doi:10.1002/path.1700510303). They compared the three tracks most used at the time — the straight edge count, the cross-sectional count and the battlement — and found the battlement gave the most accurate results. Their definition, as quoted in a University of Pretoria dissertation on the differential leukocyte count, is a “count made of three horizontal edge fields followed by two fields towards the center (so as to give three vertical fields), followed by two fields in a horizontal and then two fields in the vertical direction again”.
The same dissertation records that the battlement is the method recommended by the National Committee for Clinical Laboratory Standards (now CLSI), and that its reference differential examines each film “using the ‘battlement’ method”, with two hundred white cells counted on each film.
“Battlement” is not one path
Ask how far in the track goes and the sources disagree. The 1940 version turns two fields in from the edge; teaching notes describe an exaggerated battlement that goes a third of the way across the film, and meander methods that split the film into two or four parts and repeat that path in each. The search results for “battlement method” mix these freely, which is why they read as contradictory. They are different tracks with the same purpose.
| Track | Path | Described by |
|---|---|---|
| Straight edge | Fields adjacent to the edge only, along both edges | MacGregor et al. 1940 (via the Pretoria dissertation) |
| Cross-sectional | From one edge across the body of the film to the opposite edge, and back | MacGregor et al. 1940 (via the Pretoria dissertation) |
| Battlement | Three fields along the edge, two in, two parallel, two out, repeated | MacGregor et al. 1940; the NCCLS reference method |
| Longitudinal strip | Every cell in one strip from the head of the film to the tail | Medical technology lecture notes |
| Exaggerated battlement | From the edge inward to a third of the width, parallel to the edge, out to the edge, along it | Medical technology lecture notes |
| Two- or four-field meander | The film divided into two or four fields, each traversed as an exaggerated battlement | Medical technology lecture notes |
The lecture-note descriptions are from Lecture Notes in Medical Technology, lecture 13. Where a laboratory’s procedure names one track, that is the one to use; the comparison above is for reading the literature, not for overriding a standard operating procedure.
The mechanics the descriptions leave out
Three things decide whether a battlement count is actually a battlement count, and none of them is the shape. Move one whole field at a time, so the next field starts where the last one ended: overlapping fields count cells twice, and gaps skip them. Count every white cell in a field before moving, including the awkward one at the rim, applying the same rule each time for cells cut by the edge of the field. And stay inside the monolayer: when the inward leg would run into the thick part of the film, turn early rather than counting where the cells overlap.
The track also has an end condition the shape does not supply. On a film with a low white count a single pass may not reach a hundred cells; continue the same pattern further along the film rather than doubling back over fields already counted.
How many cells: 50, 100, 200
One hundred cells is the routine count, and it is chosen because the arithmetic is trivial: with a denominator of 100, the count of each class is already its percentage. The number is a convenience as much as a standard.
| Cells counted | Used when | Note |
|---|---|---|
| 50 | The white count is very low and 100 cells cannot be found | Reported as a 50-cell count; the imprecision is larger and should be known |
| 100 | Routine | The count of each class is its percentage |
| 200 | Markedly raised white count, or rare cells to be characterised | Results halved to give percentages |
| 400 (2 × 200) | Reference method | CLSI H20-A2: 200 cells by each of two examiners |
The classical guidance is to begin with a hundred and extend to two hundred when circumstances require it — a rare population to pin down, or a raised white count where a small shift matters. The reference standard of two examiners counting two hundred cells each is rarely run outside method comparison work, for obvious reasons of workload.
The imprecision built into the method
This is the part that is easy to work with for years without internalising, and it is not a matter of care or skill. Classifying a sample of cells is a sampling problem, and sampling has irreducible error.
The counts follow a Poisson distribution, and the coefficient of variation is approximately 1 ÷ √n. At a hundred cells that is 10%, before anyone makes a single mistake. Not 10 percentage points — 10% of the figure — but the effect on an individual class is easier to see written out.
| Cells counted | CV | A result of 6% is consistent with | A result of 60% is consistent with |
|---|---|---|---|
| 50 | 14.1% | 0% to 12.6% | 46.4% to 73.6% |
| 100 | 10.0% | 1.3% to 10.7% | 50.4% to 69.6% |
| 200 | 7.1% | 2.7% to 9.3% | 53.2% to 66.8% |
| 400 | 5.0% | 3.7% to 8.3% | 55.2% to 64.8% |
Read the hundred-cell row. A monocyte figure reported as 6% is consistent with anything from about 1% to about 11%, purely from having counted a hundred cells rather than all of them. That is the method behaving correctly. The same square-root rule sets how many cells to count on a hemocytometer, and, as in a McMaster faecal egg count, precision follows the number actually counted.
Two consequences follow, and both are practical:
- Small differences between two differentials are usually not differences. A move from 6% to 9% on successive hundred-cell counts is comfortably inside what sampling alone produces.
- Doubling the count does not halve the error. Because the error goes with √n, going from 100 to 400 cells — four times the work — halves it. That is the real reason the reference method is expensive and routine practice is not.
None of this is an argument against the manual differential. It is an argument for knowing what a percentage from it does and does not settle.
Keeping the tally without losing the field
The mechanical problem is small but real: your eyes are at the eyepieces and your hand is on the stage, so anything that requires looking away costs you your place in the field.
That is what the traditional multi-key laboratory tally counter solved — one key per class, found by touch, with a bell at a hundred. The same shape works on a keyboard: a key per class, a count that stops on its own at the target, and percentages calculated for you so no arithmetic happens at the end. What matters is that recording a cell takes no attention, because the attention belongs at the eyepiece. A reticulocyte count keeps two tallies the same way: the Miller disc method.
What no counter does is the identification. Classifying the cell is the skilled part and the whole point; the counter holds the tally and nothing more. It carries no reference ranges, makes no comment on whether a result is expected, and is not a diagnostic aid. Reference intervals are the property of the reporting laboratory (published adult ranges are compared in WBC differential normal ranges), and interpretation belongs to the clinician and the laboratory professionals responsible for the sample.
Frequently asked questions
How do you do a manual differential count?
Examine a stained film under oil immersion in the monolayer, move in a battlement pattern, and classify each white cell you meet until you reach the target count, usually 100. The counts are then reported as percentages, and may be converted to absolute counts using the total white cell count.
What is the battlement method in a differential count?
A fixed path for choosing fields on a blood film. As defined in 1940 by MacGregor, Richards and Loh, it counts three fields along the edge, two in toward the centre, two parallel to the edge and two back out, then repeats. Cells are unevenly spread on a film, so crossing between edge and centre lets the biases partly cancel.
What is the difference between the battlement and the exaggerated battlement method?
How far the track goes in. The original battlement turns two fields in from the edge before running parallel to it again. The exaggerated battlement, as described in medical technology teaching notes, goes inward to about a third of the film's width. Meander methods split the film into two or four parts and repeat that path in each. All share the aim of sampling edge and centre.
How many cells should be counted in a differential?
One hundred is routine, because each count is then already a percentage. Fifty is used when the white count is too low to find a hundred cells, and two hundred when the white count is markedly raised or a rare population needs characterising. CLSI H20-A2 sets the reference method at 200 cells by each of two examiners.
Why count in the monolayer rather than anywhere on the film?
Because the film is not uniform. Near the head the cells overlap and cannot be read; near the tail and the edges, larger cells such as monocytes and neutrophils are over-represented and lymphocytes relatively under-represented. Counting outside the monolayer introduces a bias in a known direction rather than random noise.
How accurate is a 100-cell differential?
Less than most people expect. The coefficient of variation is roughly one over the square root of the number counted, so about 10% at a hundred cells before any human error. A result of 6% is consistent with roughly 1% to 11% on sampling alone.
Is a 200-cell count twice as accurate as a 100-cell count?
No. The error falls with the square root of the number counted, so doubling the count reduces the imprecision by about 29%, not by half. Quadrupling it to 400 cells is what halves it, which is why the reference method is rarely used routinely.
Does an online differential counter interpret the result?
No, and it should not. A counter holds the tally, stops at the target and works out the percentages. Identification of each cell is the operator's, reference intervals belong to the reporting laboratory, and interpretation belongs to the clinician and the laboratory professionals responsible for the sample.
Why do the automated and manual differentials disagree?
Partly because they measure differently, and partly because a manual count of a hundred cells carries around 10% imprecision by its nature, while an analyser classifies many thousands. A small disagreement is often within what sampling alone would produce.
People also search for
More in lab and field counts
How to make a blood smear, and how to fix a bad oneDrop, angle and speed for a wedge smear, why sources give 30, 35-45 or 45 degrees, the check before staining, and each common fault traced to its cause.Published · 15 min read
McMaster faecal egg count, and where the factor comes fromThe McMaster factor derived from grams, millilitres and chambers, six published recipes compared, worked counts, and the chance that a zero count misses eggs.Published · 15 min read
Manual reticulocyte count with a Miller discThe Miller disc formula worked through, why 111 small-square cells stand for 1,000, the reticle-ratio check, counting error, and the corrected count and RPI from Hillman's 1969 data.Published · 15 min read
Platelet estimate from a blood smearThe 10-field method worked through, why the multiplier depends on your microscope's field number, the red cell ratio method that does not, and how to derive your own factor.Published · 14 min read
Mark and recapture: Lincoln-Petersen, Chapman and the confidence intervalBoth estimators worked through, which confidence interval fits your counts, the five assumptions, and what a handful of recaptures does to the answer.Published · 15 min read
How to identify white blood cells on a smear, and the look-alikesThe five white cells by size, nucleus and granules, the look-alikes that cause most misclassification, and what ICSH says each one counts as in a differential.Published · 19 min read