Frequently asked questions

What is the formula for counting cells with a hemocytometer?

Cells per mL = cells counted ÷ volume counted in mL × dilution factor. On an improved Neubauer chamber each large 1 mm square holds 0.1 µL, or 10⁻⁴ mL, so the everyday version is the average count per large square × 10⁴ × dilution. Four squares averaging 18.25 at a 100-fold dilution give 18.25 × 100 × 10⁴ = 1.8 × 10⁷ cells per mL.

Why do you multiply by 10,000 when counting cells with a hemocytometer?

Because one large square of the standard chamber is 1 mm by 1 mm with a 0.1 mm gap under the coverslip, a volume of 0.1 mm³. That is 0.1 µL, or one ten-thousandth of a millilitre. A count in one ten-thousandth of a millilitre is multiplied by 10,000 to give the count in a whole millilitre. The 10⁴ is a volume conversion, not a constant of nature.

Why does one formula use 10³ instead of 10⁴?

Because it counts ten large squares, not one. Thermo Fisher's formula C = N × D × 10³ takes N as the cells in 10 squares, which hold 1 mm³, or 1 µL, or 10⁻³ mL. Ten squares hold ten times the volume of one, so the factor is ten times smaller. Using 10³ on a single square's count gives an answer ten times too low.

Which squares do you count on a hemocytometer?

For cultured cells, usually the four large corner squares, or the four corners and the centre. The large squares are 1 mm across and suit cells that are not too dense. For small, dense cells such as red blood cells, Marienfeld's chamber guide says to count at least five of the 0.2 mm group squares inside the central square. Whatever you count, divide by the volume of exactly those squares.

Do you count cells touching the lines of a hemocytometer?

Count the cells on two adjacent sides of each square and skip those on the other two. The WHO semen manual (2010) counts cells on the lower and left lines; STEMCELL Technologies says bottom and left; ChemoMetec says top and right. Any pair works if you keep to it, because each line is then counted once, by one of the two squares it divides.

How do you calculate cell viability with trypan blue?

Viability is the unstained cells divided by all the cells counted, times 100. Live cells keep the dye out and stay clear; dead cells take it up and turn blue. With 204 clear and 22 blue cells, viability is 204 ÷ 226 × 100 = 90.3%. Thermo Fisher expects at least 95% for a healthy culture in log phase.

What is the dilution factor when you mix cells 1:1 with trypan blue?

Two. Equal volumes of cell suspension and 0.4% trypan blue, such as 10 µL of each, halve the cell concentration, so the count in the chamber is multiplied by 2 to get back to the original suspension. Strober's protocol in Current Protocols in Immunology does this explicitly, for both the viable count and the total count.

How many cells should you count on a hemocytometer?

At least 200 in total if the result matters, and more for precision. Counts follow a Poisson distribution, so the error is about 1 ÷ √N of the count: 10% at 100 cells, 7% at 200 and 5% at 400. The WHO semen manual asks for at least 200 per replicate and two replicates. STEMCELL suggests a dilution giving 50 to 100 cells per large square.

How long can cells sit in trypan blue before counting?

A few minutes at most. Strober's protocol says to count within 3 to 5 minutes of mixing, because longer incubation kills cells and lowers the viability count; Thermo Fisher says to examine the chamber immediately. The mix should also be in serum-free solution, since serum proteins take up the dye and confuse the reading.

How do you use a Fuchs-Rosenthal or Bürker chamber?

The same way, with that chamber's volumes. A Bürker large square is 1 mm² at 0.1 mm deep, 0.1 µL, exactly like a Neubauer large square, so the factor is again 10⁴. A Fuchs-Rosenthal chamber is 0.2 mm deep, so its 1 mm² squares hold 0.2 µL and one cell per square is 5,000 per mL. Always check the depth printed on the chamber.